SUB IMG

Mass Spectro Meter

 

Spare parts

Consumable Part Electronic units Prima Pro Consumable Prima Pro Electronic Units

Item NO Picture P/N Description Life Time Remark
1 PrimadB_img01 203-150-892 PRIMA ANALYZER SOURCE PRT ASSEMBLY
- The ion source is heated by a cartridge heater passing though the ion source block and the temperature is monitored by a platinum resistance thermometer (PRT) Material is made by ceramic.
if any problem, replace it.
2 PrimadB_img02 202-310-100 HEATER CARTRIDGE 2MM DIA 12V 12.5W ASI59878
- Normally the source temperature is set within the range of 140 - 200℃ (in Control Center).
if any problem, replace it.
3 PrimadB_img03 203-218-058 Cabinet Standard Filter Element
- For cabinet filtering.
Every 6 months.
4 PrimadB_img04 204-217-053 Prima Capillary Ass Mk2 (STD) M3 RMS
- Capillary assembly is typically 2.1 Metre in length. A coupling is provided to connect to the R.M.S. valve which provides a gas tight seal and minimum ‘dead volume’ to reduce purge time. After the coupling the main components are a Teflon coated stainless steel sheath with a fused silica glass liner (Flexil). The stainless steel sheath protects the liner and provides a way of heating the assembly.
if any problem, replace it.
5 PrimadB_img05 201-125-333 Turbo pump fluid reservoir TPH/U 060-071
- Working for Turbo pump. It is lubricating oil when the turbo pump working.
Every 6 months
6 PrimadB_img05 201-125-339 Turbo pump FLUID RESERVOIR FOR HiPACE 80
- Working for Turbo pump. It is lubricating oil when the turbo pump working. It looks the same as 201-125-333 model but it’s bigger.
Every 6 months
7 PrimadB_img06 203-100-668 Filament insulator plate
- It is short bar between Filament and Magnet on the Ion Source.
if any problem, replace it.
8 PrimadB_img07 203-135-305 STD Prima Thoria Filament Mesh 600S/600dB
- These have the highest emission capability and are satisfactory for the majority of mass spectrometer applications.
Yearly.
9 PrimadB_img08 201-125-990 Grade 19 rotary pump oil
- Working for Rotary pump.
Yearly.
10 PrimadB_img09 203-217-013 Face Seal (RMS)
- Working for RMS positioning system.
if any problem, replace it.
11 PrimadB_img10 201-125-283 PRIMA Inlet Probe Ferrule Glass Leak
- Checked leak on glass tube.
Yearly.
12 PrimadB_img11 203-200-841 PLATINUM PINHOLE LEAK 10 MICRON ORIFIC
- Inlet probes are normally made from glass tube around 4mm diameter and 85mm long. These glass probes are held inside a stainless steel housing which hold the probe, seals & capillary together in the correct position and configuration to maintain gas flow past through the Glass leak.
Yearly.
13 PrimadB_img12 203-100-170 PRIMA ANALYZER HALF PLATE
- Working for Ion Source.
Look for Ion burn marks or signs of contamination.
14 PrimadB_img13 203-100-225 PRIMA ANALYZER ION REPELLER PLATE
- Working for Ion Source.
Look for Ion burn marks or signs of contamination.
15 PrimadB_img14 201-110-292 ¼” 2 micron screen filter (RMS M3)
- For sample filtering on RMS.
Yearly.